Technical Specifications

 

INCUBATOR COOLING & HEATING

Temperature Range : 20°C to 60°C Accuracy : < ± 0.2°C Uniformity : < ± 1.0°C*

 

INCUBATOR HEATING

Temperature Range : 10°C above ambient to 60°C Accuracy : < ± 0.2°C Uniformity : < ± 1.0°C*
 

 

Models Cap. in liters Internal size in mm External size in mm  Tray Qty. Tray size in mm Connected load
AO/I-150 1500 1600WX600DX1700H 1760WX1210DX2060H 8 660WX580DX30H 2.5 KW
AO/I-200 2000 1600WX750DX1700H 1760WX1360DX2060H 8 660WX730DX30H 2.5 KW
AO/I-250 2500 2400WX600DX1700H 2560WX1210DX2060H 12 635WX580DX30H 2.5 KW
 AO/I-300 3000 2400WX750DX1700H 2560WX1360DX2060H 12 635WX730DX30H 2.5 KW

* Measured with high accuracy data-logger at sample storage area with standard Set Point.

Standard Chamber Features & Provisions:

  1. Racks: Heavy Duty SS304 Frame secured strongly to chamber walls for strength.
  2. Trays: Perforated types (CNC punched). Strengthened with support rails to prevent deformations.
  1. Ergonomically designed Control Panel.
  2. Ferrule printed wiring.
  3. Advanced PLC Program with built in processor-based Auto Tuning.
  4. 4.3” HMI Color LED Display.
  5. Built in Redundancy & Safety.
  6. Advanced High Accuracy “Transmitter based sensors” with output of 4-20 mA.
  1. Efficient, Reliable, Time tested & Trustworthy Refrigeration Design.
  2. Built in redundancy with PLC based Auto Change Over during power outage &/or  component failure
  3. PLC Controlled “U” Shaped SS Air Heater Banks.

 

High-Capacity Incubators:

Incubator Is a device in which incubation process is done, an enclosure conditioned to a predetermined temperature to provide and maintaining an artificially created optimal conditions for growth of microbial culture.

Incubation is process -a period of time in which temperature and other environmental variables are provided at a desired temperature level for the growth and development of microbial culture.

Incubator those works at low temperature are used for “incubation” of biochemical reactions

Industries like pharmaceutical, research laboratories, agriculture, Food & beverages requires a Cooling Incubators for their testing purposes.

Allyone’s top of the line Cooling Incubators are the most preferred choice to maintain accurate temperature & its uniformity throughout the chamber where high degree of constant temperature accuracy is required.

Incubators are mainly used for the growth & storage of bacterial cultures, Media Preparation, gentle incubation & conditioning and other purposes.

INCUBATOR HEATING

Incubator (Heating) are also known commonly as Bacteriological incubator:

This type of incubator is used majorly in laboratories mainly for the growth of bacterial cultures. it has a simple mechanism which maintains a constant set temperature as per the requirement. Accurate temperature maintained in the chamber is displayed on the display of incubator.

Most of the incubators are programmable and can be set as per the requirement. These types of incubators comes only with a heating system which maintains the temperature for growth of bacteria. These incubators are affected by temperature of surrounding environment so the prescribed set point is 10 Deg Cel above ambient temperature hence no cooling system is required or present in these chambers.

These chambers are unique offering by Allyone are offered for a niche requirement from a client, wherein the users require a bigger chamber than a stand alone chamber, but the requirement is not as big as to opt for a Walk In Chamber. Also, if there’s a space constraint these chambers are a perfect fit for such a requirement. This is a Golden Mean between the two!

The capacities we offer for our High Capacity Stand Alone Chambers are from 1500 lts (which is bigger than the max capacity of 1000 lt that we offer for a Stand Alone Chamber) to 3000 lts (which lesser than the min capacity of 4000 lt that we offer for a walk in chamber)

Some of the features are similar as of Stand Alone Type of chambers, such as it have heavy duty insulated doors : two or more-depending on the capacity, with imported hinges & locks for easy operation & inner glass door with latch to observe samples without disturbing the set condition, also the control panel with display is fixed at the top of chamber & can be easily visible & accessed.

 

 

 

 

 

 

ALLYONE ENVIRONMENTAL TECHNOLOGIES INDIA PVT LTD

Enquiry